A phantom provides a reproducible test object for a defined measurement. Choosing one starts with the question your quality assurance programme needs to answer. A CT dose phantom, an image-quality phantom and a radiotherapy motion phantom have different purposes, even when all are described as “QA phantoms”.

Start with the test, not the product photograph

Identify the equipment manufacturer and model, the measurement objective, the applicable QA protocol and the instruments already available. Your medical physicist should define the test method and acceptance criteria before a product is selected.

CT and CBCT: distinguish dose from image quality

For dose measurements, review phantom dimensions and detector access against the chosen measurement method. For image-quality checks, review the inserts and test structures needed to assess the required parameters. A product intended for one task should not be assumed to cover the other.

Also confirm whether the application is conventional CT, dental CBCT or image-guided radiotherapy. The acquisition method, field of view and evaluation protocol matter. Published CT/CBCT phantom research illustrates why CT image-quality tolerances cannot simply be transferred to CBCT.

Explore diagnostic radiology QA phantoms →

Radiotherapy: define the verification workflow

Specify whether you need an anatomical setup, a film measurement insert, a positioning check or a moving target for a respiratory-motion test. For a multi-step verification, describe the steps to be tested, from imaging and planning through delivery and measurement.

Check the tissue-equivalent materials, fiducial arrangement, film slots or detector cavities, positioning accessories and any required motion system. A static phantom should not be assumed to provide motion testing without an appropriate motion assembly.

Our modular radiosurgery range includes head-and-neck, AQA, film-target, dome, thorax and respiratory configurations. The required modules should be selected for the intended test. Explore radiotherapy QA phantoms →

Nuclear medicine: separate image quality from registration

For PET, PET/CT and SPECT/CT, distinguish image-quality measurements from alignment between imaging modalities. Review whether the proposed phantom has the required internal structures, filling arrangements and markers for the selected procedure.

The EANM guidelines for PET-CT and PET-MR routine quality control describe different QC tasks, including CT checks and multimodality alignment. A phantom suitable for one task does not automatically meet the requirements of every other task.

Explore nuclear medicine QA phantoms →

Check what the quotation includes

Confirm each item explicitly. A phantom quotation should not be assumed to include a detector, electrometer, film scanner or analysis software.

What to send us for a recommendation

Please share your system model, planned tests, protocol or tender specification, required dimensions and existing measurement equipment. If replacing a phantom, include photographs and the original configuration. We can then identify suitable options and list any deviations before quoting.

For wider background, see the IAEA Handbook of Basic Quality Control Tests for Diagnostic Radiology. References guide test selection; they do not certify that a particular supplied product meets a protocol.

Discuss your requirements

Send your configuration and quantity to sales@rgodosimetry.com for product information and a quotation.